Echometer Dynamometer Transducer Resistance Values

All values are +/- 25%

Four-inch 30,000 lb Load Cells and 
Five-inch 50,000 lb Load Cells

Pins Values Circuit
     
Pins A to B 730 Ohms Load
Pins A to F 730 Ohms Load
Pins D to B 730 Ohms Load
Pins D to F 730 Ohms Load
Pins A to C 2.2 K Ohms Accel.
Pins A to E 2.2 K Ohms Accel.
Pins D to C 2.2 K Ohms Accel.
Pins D to E 2.2 K Ohms Accel.
     
Pins A to D 840 Ohms Load and Accel.
Pins B to F 1.1 K Ohms Load bridge output
Pins C to E 4.3 K Ohms Accel. bridge output

Load Cells with serial numbers HT137 through HT508 had pin "D" grounded to case when manufactured. The grounding wire from pin "D" to case was removed if the transducer has been serviced by Echometer Company.